溶氧電(dian)極(ji)OXYFERM FDA采(cai)用極(ji)譜(pu)式(shi)電(dian)極(ji),陽(yang)電(dian)極(ji)為(wei)Ag/AgCl、陰電(dian)極(ji)為(wei)鉑金(Pt)組(zu)成(cheng),兩者(zhe)之間充滿特殊(shu)成(cheng)份的是電(dian)解液。由(you)矽(gui)橡(xiang)膠(jiao)滲(shen)透膜包(bao)裹(guo)於(yu)電(dian)極(ji)四(si)周。測(ce)量時(shi),電(dian)極(ji)間加(jia)上675mV的極(ji)化電(dian)壓(ya),氧(yang)滲透過(guo)隔膜在(zai)陰(yin)極(ji)消(xiao)耗,同(tong)時(shi)等量的氧在(zai)陽(yang)極(ji)產(chan)生,這(zhe)個(ge)動態過(guo)程進行到(dao)兩邊(bian)的氧分(fen)壓(ya)相同(tong)時(shi)達(da)到(dao)平衡(heng).此(ci)時(shi)兩(liang)電(dian)極(ji)間的電(dian)流(liu)與(yu)氧(yang)分壓成正比,二次表(biao)檢測到(dao)此(ci)電(dian)流(liu),再(zai)經(jing)過(guo)壹系(xi)列(lie)變換(huan),得(de)到(dao)氧濃(nong)度(du)和(he)氧(yang)含量.同(tong)時(shi),NTC(負(fu)溫度(du)系(xi)數(shu)熱(re)敏(min)電(dian)阻(zu))檢測被(bei)測(ce)量液的溫度(du),二(er)次表(biao)采樣(yang)後(hou)進(jin)行(xing)溫(wen)度(du)補(bu)償(chang),將氧濃(nong)度(du)或(huo)氧(yang)含量折算(suan)成25℃時(shi)的值(zhi)。
1、用純水或(huo)蒸(zheng)餾水配置飽和亞(ya)硫(liu)酸鈉溶液,水溫(wen) 25 度(du)左(zuo)右(you)或略高(gao);
2、溶氧電(dian)極(ji)在(zai)空氣中校(xiao)正(zheng) 100%,電(dian)極(ji)斜(xie)率(lv)正(zheng)常;
3、將電(dian)極(ji)插(cha)入亞硫(liu)酸鈉飽和溶液中,輕(qing)輕(qing)晃(huang)動電(dian)極(ji);
4、極(ji)譜(pu)式(shi)溶氧電(dian)極(ji) 2 分(fen)鐘測(ce)量數據降(jiang)到(dao) 10% 左(zuo)右(you),5 分鐘測(ce)量數據降(jiang)到(dao) 1% 左(zuo)右(you),10 分鐘測(ce)量數據降(jiang)到(dao) 1% 以下或(huo)接近於(yu) 0;
5、光(guang)學(xue)溶氧電(dian)極(ji)在(zai)亞硫(liu)酸鈉飽和溶液中速(su)度(du)略慢,10 分鐘測(ce)量數據在 1% 左(zuo)右(you);
6、將電(dian)極(ji)從(cong)飽和亞(ya)硫(liu)酸鈉溶液中拿出來置於(yu)空氣中,用吸(xi)水紙(zhi)吸(xi)幹電(dian)極(ji)頭前端(duan)的水珠(zhu),20 秒(miao)左(zuo)右(you)測量數據恢(hui)復到(dao) 100%;
7、如果(guo)滿足以上測量過(guo)程,說明電(dian)極(ji)、電(dian)纜(lan)、變(bian)送(song)器(qi)均正(zheng)常;
8、如果(guo)沒(mei)有亞硫(liu)酸鈉飽和溶液,可以用聽裝(zhuang)的可口可樂(le)或生產(chan)日期稍長的聽裝(zhuang)啤(pi)酒(jiu)來測試(shi),效果(guo)基(ji)本(ben)壹(yi)致。