溶(rong)氧(yang)電(dian)極(ji)OXYFERM FDA故障的解(jie)決方(fang)案
發(fa)布(bu)時(shi)間: 2018-08-27 點擊次數: 1594次(ci)
溶(rong)氧(yang)電(dian)極(ji)OXYFERM FDA故障的解(jie)決方(fang)案
溶(rong)氧(yang)電(dian)極(ji)OXYFERM FDA采(cai)用(yong)極譜式電(dian)極(ji),陽(yang)電(dian)極(ji)為(wei)Ag/AgCl、陰電(dian)極(ji)為(wei)鉑金(jin)(Pt)組成,兩(liang)者之(zhi)間充滿特殊成份(fen)的是電(dian)解(jie)液(ye)。由(you)矽橡(xiang)膠滲(shen)透膜(mo)包(bao)裹(guo)於(yu)電(dian)極(ji)四(si)周(zhou)。測(ce)量時(shi),電(dian)極(ji)間加上(shang)675mV的(de)極(ji)化(hua)電(dian)壓(ya),氧滲(shen)透過(guo)隔膜(mo)在(zai)陰極消(xiao)耗,同(tong)時(shi)等(deng)量的氧在(zai)陽(yang)極產生(sheng),這(zhe)個(ge)動(dong)態過(guo)程(cheng)進行(xing)到兩(liang)邊(bian)的氧(yang)分(fen)壓(ya)相(xiang)同(tong)時(shi)達到平(ping)衡.此(ci)時(shi)兩(liang)電(dian)極(ji)間的電(dian)流(liu)與氧(yang)分(fen)壓(ya)成正(zheng)比(bi),二(er)次表檢測(ce)到(dao)此(ci)電(dian)流(liu),再經過壹(yi)系列變(bian)換(huan),得到(dao)氧濃度和氧(yang)含量.同(tong)時(shi),NTC(負(fu)溫度(du)系(xi)數(shu)熱敏(min)電(dian)阻(zu))檢(jian)測(ce)被測(ce)量液(ye)的(de)溫(wen)度(du),二(er)次表采樣(yang)後(hou)進行(xing)溫度(du)補償,將氧(yang)濃(nong)度或氧(yang)含(han)量折算(suan)成25℃時(shi)的(de)值(zhi)。
溶(rong)氧(yang)電(dian)極(ji)OXYFERM FDA故障處理(li)
使用(yong)時(shi),校正(zheng)或測(ce)量前要預(yu)熱至(zhi)少15-30分(fen)鐘。
為(wei)確保膜(mo)的(de)電(dian)解(jie)液(ye)內沒有氣(qi)泡(pao),ASI膜(mo)帽(mao)在(zai)設計上(shang)要(yao)求(qiu)在(zai)裝(zhuang)上(shang)膜(mo)頭(tou)時(shi)要(yao)排(pai)除(chu)掉(diao)所有液(ye)腔(qiang)內的空氣(qi)。
膜(mo)表(biao)面上(shang)不(bu)能(neng)留有任(ren)何氣(qi)泡(pao),否(fou)則它(ta)會將氣(qi)泡(pao)當(dang)作(zuo)氧(yang)飽和樣(yang)品(pin)進行(xing)讀(du)數。
即使使用(yong)的是帶有自(zi)動(dong)溫度(du)補償的儀表,也(ye)要(yao)在(zai)接(jie)近(jin)樣(yang)品(pin)溶(rong)液(ye)的(de)溫(wen)度(du)下校正(zheng)電(dian)極(ji)。
電(dian)極(ji)應(ying)在(zai)空氣(qi)中(zhong)校正(zheng),以空氣(qi)作(zuo)為100%的(de)飽和溶(rong)解(jie)氧標(biao)準(zhun)點。
由於(yu)電(dian)極(ji)對(dui)氧的(de)消(xiao)耗,探頭(tou)表面氧的(de)濃度(du)會瞬(shun)間降(jiang)低,因此(ci)測(ce)量時(shi)要(yao)對(dui)溶(rong)液(ye)進行(xing)攪拌(ban),這非常重(zhong)要(yao)。
如(ru)膜(mo)已破損(sun)則要進行(xing)更換(huan)。