便攜(xie)式(shi)熒光(guang)法(fa)溶(rong)氧(yang)儀9030M基(ji)於熒光(guang)猝熄原理,藍(lan)光(guang)照射到(dao)熒光(guang)物(wu)質(zhi)上(shang)使熒光(guang)物(wu)質(zhi)激(ji)發(fa)並(bing)發(fa)出紅(hong)光,由(you)於氧(yang)分子(zi)可以帶走能(neng)量(liang)(猝熄效(xiao)應),所(suo)以(yi)激發的紅(hong)光的時(shi)間(jian)和強度與氧(yang)分子(zi)的濃度成反比。通(tong)過(guo)測(ce)量(liang)激發紅(hong)光與(yu)參(can)比(bi)光的相位(wei)差(cha),並(bing)與(yu)內部標(biao)定(ding)值(zhi)對比(bi),從而可計算出氧(yang)分子(zi)的濃度。
更(geng)換電(dian)極或(huo)電(dian)解液(ye)後(hou)得用(yong)碘量(liang)法來檢(jian)查(zha)校(xiao)準(zhun)。
檢(jian)查(zha)方(fang)法(fa):通(tong)過(guo)測(ce)定(ding)壹系(xi)列不(bu)同(tong)濃度蒸餾水樣(yang)品中溶(rong)解氧(yang)的濃度來檢(jian)查(zha)儀(yi)器的線性。向3~4 個(ge)250毫升*充滿蒸(zheng)餾水的細口(kou)瓶中緩(huan)緩通入(ru)氮(dan)氣泡,去除水中氧(yang)氣,用探頭時(shi)刻(ke)測(ce)量(liang)剩余(yu)的溶(rong)解氧(yang)含量(liang),直(zhi)到獲(huo)得所(suo)需(xu)溶(rong)解氧(yang)的近(jin)似(si)濃(nong)度,然後(hou)立刻(ke)停止通(tong)氮(dan)氣,用GB 7489 測(ce)定(ding)水中準(zhun)確的溶(rong)解氧(yang)濃度。若探頭法測(ce)定(ding)的溶(rong)解氧(yang)濃度值(zhi)與碘(dian)量法(fa)在(zai)顯(xian)著性水平(ping)為5%時(shi)無顯(xian)著性差異(yi),則認為探頭的響(xiang)應呈(cheng)線性。
便攜(xie)式(shi)熒光(guang)法(fa)溶(rong)氧(yang)儀9030M電(dian)極可(ke)以校(xiao)準(zhun),看(kan)樓主(zhu)有(you)多(duo)高的要求(qiu)。根(gen)據樓主(zhu)對於準(zhun)確度要求(qiu)的不同可以(yi)分(fen)為單點校(xiao)正(zheng),兩(liang)點校(xiao)正(zheng)和多(duo)點校(xiao)正(zheng)。
單(dan)點校(xiao)正(zheng):將蒸(zheng)餾水緩(huan)速(su)通空氣進行(xing)飽(bao)和(10分鐘左右),記錄(lu)水溫(wen),壓(ya)力視為壹個(ge)大氣壓(ya),將溶(rong)解氧(yang)電極(ji)插(cha)入(ru)測(ce)定(ding)溶(rong)解氧(yang)含量(liang),記錄(lu)。根據溫(wen)度查物化表計算在(zai)此溫(wen)度下水中溶(rong)解氧(yang)的飽和濃度,飽和濃度除以溶(rong)氧(yang)電極(ji)測(ce)定(ding)值(zhi)即為校(xiao)正(zheng)因(yin)子(zi)(K),測(ce)定(ding)樣品中溶(rong)解氧(yang)的值(zhi),乘以(yi)校(xiao)正(zheng)因(yin)子(zi)即為實際(ji)值(zhi)。溶(rong)解電極(ji)每天至少(shao)校(xiao)正(zheng)壹次(ci),如(ru)果樣(yang)品溫(wen)度變(bian)化(hua)較(jiao)大,則以(yi)需(xu)要經常(chang)校(xiao)正(zheng)。
雙(shuang)點校(xiao)正(zheng):溶(rong)解氧(yang)飽和溶(rong)液(ye)的制取同(tong)上(shang)。另需(xu)制取溶(rong)解氧(yang)零點溶(rong)液(ye)。可(ke)以通過(guo)向水(shui)中通(tong)高純N2,He的辦法(fa)制取,He的效(xiao)果比(bi)N2要好壹些(xie)。用(yong)溶(rong)氧(yang)電極(ji)測(ce)定(ding)零點的溶(rong)氧(yang)值(zhi),與剛(gang)才(cai)測(ce)定(ding)的飽和溶(rong)液(ye)的溶(rong)氧(yang)值(zhi)壹起(qi)計算計算校(xiao)正(zheng)因(yin)子(zi)。
多(duo)點校(xiao)正(zheng):配(pei)置不同(tong)濃(nong)度系(xi)列(3-5個(ge))的溶(rong)解氧(yang)溶(rong)液(ye),分(fen)別利(li)用(yong)電極測(ce)定(ding)溶(rong)解氧(yang)值(zhi),同時(shi)利(li)用(yong)碘量滴(di)定(ding)法測(ce)定(ding)上(shang)述(shu)3-5個(ge)溶(rong)液(ye)的溶(rong)解氧(yang)值(zhi),兩(liang)個(ge)測(ce)定(ding)系(xi)列線性回歸,斜(xie)率即為校(xiao)正(zheng)因(yin)子(zi)。